รายละเอียด
|
AN9651TH (F)
|
การทดสอบกราว์บอน (Ground Bond Test)
|
Empty Load Output Voltage
|
<12V
|
Output Current Setting Range
|
AC3A~30A 50Hz/60Hz,1A/step
|
Output Current Setting Accuracy
|
±(2% of setting+0.2A)
|
Upper Limit of Resistance
|
3A≤output current≤10A, 10~600mΩ; 10A
|
Ground Bond Resolution/ Measurement Range/ Accuracy
|
1mΩ, 10~600mΩ, ±(2% of reading +2mΩ)
|
Compensation Resistance Measurement Range/Accuracy
|
0~200mΩ 1mΩ
|
Test Time Setting Range
|
0.5~999.9s 0.1s/step
|
การทดสอบความเป็นฉนวน (Insulation Resistance Test)
|
Rated Output Load/Current
|
1VA (1000V/1mA) 1mA
|
Output Voltage Range/Adjustable
|
DC:500V/1000V <1% (Empty load to full load)
|
Output Voltage Ripple
|
Pure resistance load, 1000V/1mA, <3%
|
Insulation Resistance Range
|
1.0~2000MΩ
|
Insulation Resistance Lower Limit Setting Range/ Resolution
|
1.0~99.9 MΩ, 0.1MΩ/step, 100~2000 MΩ 1 MΩ/step
|
Insulation Resistance Accuracy
|
1~200 MΩ: ±(2% of reading+2counts) , 200~2000 MΩ: ±(5% of reading+5counts)
|
Ramp-up Time Range/Resolution
|
0.1~999.9s 0.1s
|
Delay Judgment Time
|
0.3~10.0s (Count from ramp-up )
|
Test Time Setting Range
|
0.5~999.9s 0.1s/step
|
การทดสอบความต้านทานแรงดันไฟสูง AC (AC Hipot Test)
|
Rated Output Load
|
500VA (AC:5000V/100mA)
|
Output Voltage Range /Frequency/Adjustable Rate
|
500~5000V, 50Hz/60Hz, <2%(Empty load to full load)
|
Output Voltage Setting Range/Accuracy
|
Cold state AC:500~5000V, Hot state AC:500~2500V, 5V/step, ±(2.5% of setting+5V)
|
Leakage Current Preset Alarm
|
Upper limit: 0.10~99.99mA; Lower limit: 0.00~99.99mA , 0.01mA/step
|
Compensation Current Setting Range/ Resolution
|
0.00 ~5.00mA, 0.01mA
|
Output Initial Voltage Range
|
(0%~99%) of test voltage, from 0V to initial voltage need 200ms
|
Leakage Current Measurement Range/Resolution/Accuracy
|
0.10~99.99 mA, 0.01mA, ±(2% of reading +2counts)
|
Test Time/Ramp-Up/Ramp-Down
|
0.5~999.9s, 0.1~999.9s, 0.0~999.9s(After test pass)
|
การทดสอบความต้านทานแรงดันไฟสูง DC (DC Hipot Test)
|
Rated Output Load
|
60VA (AC:6000V/10mA)
|
Output Voltage Range /Frequency/Adjustable Rate
|
500~6000V, <2%(Empty load to full load)
|
Output Voltage Setting Range/Accuracy
|
Cold state AC:500~6000V, Hot state AC:500~2500V, 5V/step, ±(2.5% of setting+5V)
|
Leakage Current Preset Alarm
|
Upper limit: 50~9999μA; Lower limit: 0~9999μA , 1μA
|
Compensation Current Setting Range/ Resolution
|
0 ~5000μA, 1μA
|
Output Initial Voltage Range
|
(0~99%) of test voltage, from 0V to initial voltage need 200ms
|
Leakage Current Measurement Range/Resolution/Accuracy
|
50~9999μA , 1μA, ±(2% of reading +2counts)
|
Test Time/Ramp-Up/Ramp-Down
|
0.5~999.9s, 0.1~999.9s, 0.3~10.0s(Count from ramp-up)
|
การทดสอบกระแสไฟรั่วไหล (Leakage Current Test)
|
Output Voltage Range
|
1.06 times of input voltage (output by isolated transformer)
|
Leakage Current Range/Resolution
|
Upper limit: 50~9999μA 1μA
|
Compensation Current Setting Range/ Resolution
|
0~500μA 1μA
|
Test Time Range/Resolution
|
1.0 ~999.9s 0.1s
|
Leakage Type
|
dynamic, static
|
Three Phase Dynamic Mode
|
A mode: ABC-AB-BC-CA (standard) B mode: ABC-AB-ABC-BC-ABC-CA (customized)
|
Output Voltage Measurement Range/Accuracy
|
AC 50~500.0V 0.1V, ±(0.5% of reading+2counts)
|
Leakage Current Measurement Range/Accuracy
|
50~9999μA 1μA DC~10kHz: ±(2% of reading+5counts) 10kHz~1MHz: ±(5% of reading+5counts)
|
Leakage Current Accuracy
|
50.0~2000μA is ±(5% of reading);2000μA or above is ±(2.5% of reading)
|
Leakage Power
|
6kVA ,isolated transformer capacity is designed according to the rated voltage 220V
|
Time Setting Range
|
1~300s
|
Measurement Mode
|
dynamic, static
|
การทดสอบค่าพลังงานไฟฟ้า (Power Test)
|
ย่านแรงดัน.เอาท์พุท
|
1.00 times of input voltage (output by isolated transformer)
|
Power Setting Range/Resolution
|
Upper limit, Single phase:20~6000W Three phase:60~20000W Lower limit, Single phase:0~6000W Three phase:0~20000W 1W/step
|
Test Time Range/Resolution
|
0.5 ~999.9s 0.1s
|
Voltage Measurement Range and Accuracy
|
AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)
|
Current Measurement Range and Accuracy
|
AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts)
|
Active Power Measurement Range and Accuracy
|
20.0~999.9W; 1.000~9.999kW; 10.00~20.00kW 45~65Hz: ±(0.5% of reading +2counts)
|
Power Factor Range and Accuracy
|
0.100~1.000, ±(2% of reading +2counts)
|
การทดสอบช่วงริ่มต้นทำงานที่แรงดันไฟต่ำ (Low Voltage Start Test)
|
Output Voltage Range
|
0.85 times of input voltage (output by isolated transformer)
|
Test Time Range/Resolution
|
0.5 ~999.9s 0.1s
|
Current Setting Range/Resolution
|
Upper limit: 0.30 ~30.00A Lower limit: 0.00 ~30.00A0.01A
|
Voltage Measurement Range and Accuracy
|
AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)
|
Current Measurement Range and Accuracy
|
AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts)
|
Display Mode
|
10.4 Inch, 640×480, TFT LCD display
|
Isolated Transformer Capacity
|
Three Phase:20kVA, Single Phase: 6kVA
|
Over Current Protection
|
Power, Start, Leakage Current, Phase Current35A, over 5s protection
|
DUT Type Setting
|
Single Phase, Three phase Three line, Three phase Four line
|
Grounding Mode
|
Load Grounding, Load Floating
|
DUT Error Processing
|
Voice and light alarm
|
Outer Alarm Light
|
Red alarm light
|
PLC Output Interface
|
Each test function have one
|
Interface Function
|
RS232, Outer Printer, Remote
|
Optional Interface
|
RS485, GPIB, Internet Card, Inner Printer
|
Remote control Interface
|
Foot Pedal, three-color alarm light(on-off signal output, PRINTER, 97 COM interface, Mini-Printer function optional)
|
Test Group
|
15 groups, 8 steps / group
|
Buzzer Alarm Voice
|
Can turn on ,Can cut
|